Z-2149-2014Class II
SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309044 - 100 test For in vitro diagnostic use in...
Siemens Healthcare Diagnostics Inc. / initiated 2014-06-23 / Terminated / root cause: manufacturing process
Siemens Healthcare Diagnostics, Inc began a recall of SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309044 - 100 test For in vitro diagnostic use in the quantitative determination of B-type... on . FDA classified it as Class II and gives the reason as "Complaints of failed calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2149-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309044 - 100 test For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur CP System.
- Product code
- NBC Test, Natriuretic Peptide
- Reason for recall
- Complaints of failed calibration.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Siemens sent an Urgent Medical Device Correction (UMDC) letter to all affected Siemens Healthcare Diagnostics customers in the United States on June 23, 2014 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on June 23, 2014 via hard copy, email, and/or fax per each country's local regulations. These notices inform the customer of the potential for low calibrator deviation failures when using ADVIA Centaur Systems BNP kit lots ending in 172 on the ADVIA Centaur CP due to an error on the master curve card. Siemens informs customers if a successful calibration is achieved, the results obtained are valid. If customers are unable to obtain a successful calibration on the ADVIA Centaur CP, they should switch to an alternative lot of BNP assay. The notices should be reviewed with the facility's laboratory or medical director. Fax-back forms will be used to confirm that the customers received the UMDC. For those customers that do not return a fax-back form, a Siemens representative will follow up with those customers to ensure they received the Urgent Medical Device Correction and completed the required actions. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 877-229-3711.
- Quantity in commerce
- 1,191 kits
- Distribution
- Worldwide Distribution - USA (nationwide including Puerto Rico) and Internationally to Canada, Albania, Argentina, Australia, Bosnia Herzeg., Brazil, Chile, Colombia, Egypt, Georgia, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Japan, Kazakhstan, Lebanon, Malaysia, New Caledonia, Oman, P.R. China, Pakistan, Philippines, Qatar, Rep. of Yemen, Republic Korea, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, Taiwan, Thailand, U.A.E., Ukraine, Venezuela, Vietnam, Austria, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece,Hungary,Iceland, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Poland, Portugal, Romania,Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Product codes and lots
- Lot# 38479172 and 39181172
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2149-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2150-2014 | Class II | Terminated | Voluntary: Firm initiated |