510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2092-2014Class II

Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior...

Alphatec Spine, Inc. / initiated 2014-06-26 / Terminated

Alphatec Spine, Inc. began a recall of Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization System, a spinal fixation system intended to... on . FDA classified it as Class II and gives the reason as "Alphatec Spine is recalling the 2.1mm Fixed Depth Drill 10mm because they received a report that the instrument malfunctioned and golden stop slid from its fixed position during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2092-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alphatec Spine, Inc.
Product
Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization System, a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Alphatec Spine is recalling the 2.1mm Fixed Depth Drill 10mm because they received a report that the instrument malfunctioned and golden stop slid from its fixed position during use.
FDA root cause
Unknown/Undetermined by firm
Action
Alphatec Spine, Inc. sent an Urgent Medical Device Notification letter dated June 20, 2014, June 26, 2014, to all affected customers. Alphatec Spine received a report of the 10mm fixed depth drill from this lot which malfunctioned and golden stop slid from its fixed position during use. The letter informs the customers of the problems identifed and the actions to be taken. Customers are instructed to confirm that they received the notification letter by completing the customer response form. Completed forms are to be sent to Alphatec Spine at 5818 El Camino Rd, Carlsbad, CA 92008. Customers may also return forms by fax at (760) 431-0289. Customers with questions were instructed to contact Alphatec Spine's Customer Service Department on how to return the affected product. For questions regarding this recall call 760-431-9286.
Quantity in commerce
14 units total (12 units in US)
Distribution
Nationwide Distribution including CT, NC, AL, WV, IL, AR, and NY.
Product codes and lots
Lot No. 7117902

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2092-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2092-2014Class IITerminatedVoluntary: Firm initiated