Z-2153-2014Class II

Panthera-Puncher 9 . Panthera-Puncher 9 is a sample transfer instrument for professional use. The sample, a...

Revvity Health Sciences, Inc. / initiated 2014-07-01 / Terminated / root cause: manufacturing process

PerkinElmer Health Sciences, Inc. began a recall of Panthera-Puncher 9 . Panthera-Puncher 9 is a sample transfer instrument for professional use. The sample, a dried blood disk, is extracted from a larger area... on . FDA classified it as Class II and gives the reason as "Risk of incorrect results from samples prepared with Panthera-Puncher 9". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2153-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Revvity Health Sciences, Inc.
Product
Panthera-Puncher 9 . Panthera-Puncher 9 is a sample transfer instrument for professional use. The sample, a dried blood disk, is extracted from a larger area of the sample (dried blood spot) by punching the disk of the specified size to a sample collection vessel, e.g. a 96-well microtiter plate. The identification of the sample punched to a specific well of the plate is stored and carried on to the next instrument or the data is transferred to the LIMS.
Product code
N/A
Reason for recall
Risk of incorrect results from samples prepared with Panthera-Puncher 9
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Perkin Elmer sent an Urgent Medical Device Recall letter dated July 1, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The Recall Letter delivered to the end customer by email facilitated through our Perkin Elmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: "Removing the plates from the Panthera-Puncher 9 when not actively punching and processing samples in the instrument. "Opening the plate track hatch if plates must reside in the Panthera-Puncher 9 while the instrument is idle or in sleep mode. This will interrupt power to the ionizer and eliminate any ionizer impact on the plate wells. For further information, please contact your local Perkin Eimer representative or [email protected]. For further information please call (781) 663-6900.
Quantity in commerce
144 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of : Canada, Australia, Brazil, China, Denmark, Great Britain, Italy, France, Mexico, Qatar and Korea.
Product codes and lots
Serial Numbers: 20810001 through 20810148

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2153-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2153-2014Class IITerminatedVoluntary: Firm initiated