Z-2036-2014Class III

Reliance 1227 Cart and Utensil Washer/Disinfector

STERIS Corporation / initiated 2014-06-30 / Terminated / root cause: design

Steris Corporation began a recall of Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the... on . FDA classified it as Class III and gives the reason as "STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2036-2014
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.
Product code
MEC Disinfector, Medical Devices
Reason for recall
STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.
Root cause tag
design
FDA root cause
Device Design
Action
Customers will be notified via onsite visits by STERIS Field Service Technicians to each affected Customer site. These visits will be documented and monitored until completion.
Quantity in commerce
37
Distribution
US Nationwide Distribution - in the states of AL, AR, CA, DE, FL, GA, IL, IA, LA, MA, MI, NJ, NY, OH, OK, OR, SC, TX, VA. and in the country of Canada.
Product codes and lots
Serial Number Range: 3632511020 - 3613214010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2036-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2036-2014Class IIITerminatedVoluntary: Firm initiated