Reliance 1227 Cart and Utensil Washer/Disinfector
Steris Corporation began a recall of Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the... on . FDA classified it as Class III and gives the reason as "STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2036-2014
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.
- Product code
- MEC Disinfector, Medical Devices
- Reason for recall
- STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Customers will be notified via onsite visits by STERIS Field Service Technicians to each affected Customer site. These visits will be documented and monitored until completion.
- Quantity in commerce
- 37
- Distribution
- US Nationwide Distribution - in the states of AL, AR, CA, DE, FL, GA, IL, IA, LA, MA, MI, NJ, NY, OH, OK, OR, SC, TX, VA. and in the country of Canada.
- Product codes and lots
- Serial Number Range: 3632511020 - 3613214010