Z-2147-2014Class II

Boston Scientific ENDOTAK RELIANCE SG, transvenous defibrillation lead, Models 0180 & 0292, Sterile EO

Boston Scientific Corporation / initiated 2014-06-19 / Terminated / root cause: software

Boston Scientific CRM Corp began a recall of Boston Scientific ENDOTAK RELIANCE SG, transvenous defibrillation lead, Models 0180 & 0292, Sterile EO. Product Usage: The ENDOTAK RELIANCE leads provide... on . FDA classified it as Class II and gives the reason as "A review of manufacturing test records for Boston Scientific ENDOTAK RELIANCE implantable leads revealed a suspected test data recording error. Specifically, some test results were recorded as "failed" without any other...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2147-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific ENDOTAK RELIANCE SG, transvenous defibrillation lead, Models 0180 & 0292, Sterile EO. Product Usage: The ENDOTAK RELIANCE leads provide pacing and rate-sensing and deliver cardioversion and defibrillation shocks for automatic implantable cardioverter defibrillator (AICD) systems.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
A review of manufacturing test records for Boston Scientific ENDOTAK RELIANCE implantable leads revealed a suspected test data recording error. Specifically, some test results were recorded as "failed" without any other indication of failure. There are no reported injuries from the devices.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Sales representatives hand delivered a Boston Scientific "Medical Device Retrieval" Letter dated June 19, 2014 or June 20, 2014 when they were retrieving the devices from the hospital shelves. The letter was addressed to Hospital Administrator. The letter described the problem, product being recalled and the retrieval of the device by Boston scientific sales representative. For further questions they can contact their local sales representative or Boston Scientific International Technical Services. Physicians were contacted via telephone starting on July 6, 2014. A Boston Scientific letter dated July 8, 2014 followed the telephone conversation. For further information physicians can contact their local Boston Scientific representative or Technical Services at 1-800-227-3422.
Quantity in commerce
21
Distribution
Worldwide Distribution -US (nationwide) in the states of OH, MN and country of France
Product codes and lots
Model 0180 s/n 310735 & 310892. Model 0292 s/n 130450, & 309339.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2147-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2147-2014Class IITerminatedVoluntary: Firm initiated