Artis zee and Artis zeego systems. x-ray, angiographic system
Siemens Medical Solutions USA, Inc began a recall of Artis zee and Artis zeego systems. x-ray, angiographic system on . FDA classified it as Class II and gives the reason as "There is a potential problem with Artis zee and Artis zeego systems running software version VC21B and being used in conjunction with the Large Display, in that under certain circumstances, the release of radiation can...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2569-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zee and Artis zeego systems. x-ray, angiographic system
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- There is a potential problem with Artis zee and Artis zeego systems running software version VC21B and being used in conjunction with the Large Display, in that under certain circumstances, the release of radiation can become blocked unnecessarily.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent an "Important Customer Safety Notice" dated August 21, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please forward this Safety information to other organizations that could be affected by this measure. If the device has been sold and therefore no longer in your possession, please forward this safety notice to the new owner. We would also request that you inform us of the identity of the device's new owner where possible. Please observe this Customer Safety Notice and comply with the corresponding measures until the update has been fully completed. For further questions please call (610) 219-6300.
- Quantity in commerce
- 206
- Distribution
- Nationwide Distribution
- Product codes and lots
- running software version VC21B 10094135, 10094137, 10094141, 10280959 with multiple serial numbers .