Z-0182-2015Class II

Philips Medical Systems Allura Xper FD20C with software version R8.2.O System Code: 722035. The Allura Xper...

Philips North America LLC / initiated 2014-07-08 / Terminated / root cause: software

Philips Medical Systems, Inc. began a recall of Philips Medical Systems Allura Xper FD20C with software version R8.2.O System Code: 722035. The Allura Xper FD series are intended for use on human patients to... on . FDA classified it as Class II and gives the reason as "Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0182-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Medical Systems Allura Xper FD20C with software version R8.2.O System Code: 722035. The Allura Xper FD series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
Root cause tag
software
FDA root cause
Software design
Action
Philips Medical Systems issued certified letters on 7/8/14 and receipt returned for tracking purposes. A mandatory Field Change Order with reference FC072200253 is being released that requires Philips field service engineers to install Software release R8.2.0.3. If you need any further information or support concerning this issue, contact your local Philips representative: SUPPORT 1-800-722-9377.
Quantity in commerce
2 units
Distribution
Distributed in the states of IL, and TN, and the countries of Argentina , Australia, Belgium, Denmark, Egypt, France, Germany, Hong Kong, Italy, Japan, Korea, Republic of Netherlands, Norway, Saudi Arabia, Switzerland, United Kingdom, and Viet Nam.
Product codes and lots
Allura Xper Software version R8.2.0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0182-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0181-2015Class IITerminatedVoluntary: Firm initiated
Z-0182-2015Class IITerminatedVoluntary: Firm initiated