Z-2177-2014Class II

Kodak DirectView DR 7500 Diagnostic X-Ray System

Eastman Kodak Company / initiated 2014-07-09 / Terminated / root cause: design

Carestream Health, Inc. began a recall of Kodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently installed diagnostic x-ray system composed of 4 main... on . FDA classified it as Class II and gives the reason as "An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering. Inspection...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2177-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Eastman Kodak Company
Product
Kodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently installed diagnostic x-ray system composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DR 7500 System is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Product code
KPR System, X-Ray, Stationary
Reason for recall
An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering. Inspection of these parts revealed that the tab on the lock washer was not bent into the lock nut recess and because the Beta Assembly detached from the OTC telescope, the lock nut was not sufficiently tightened.
Root cause tag
design
FDA root cause
Device Design
Action
Carestream will notify all DR 7500 customers of the issue and schedule a Field Engineer to inspect and repair the system if required. Specific instructions (Service Modification) for the inspection and repair will followed by the Carestream Field Engineer in order to complete the correction. Documented evidence of the Field Engineer visit will be noted in the Carestream service engineering record for each affected system. Carestream estimates that the plan of correction will be completed within 2-4 weeks
Quantity in commerce
589 systems
Distribution
Worldwide Distribution - US Nationwide and the countries of (United Kingdom, United Arab Emirates, Turkey, Taiwan, Switzerland, Sweden, Spain, South Africa, Portugal, Poland, Oman, New Zealand, Netherlands, Kuwait, Italy, Israel, Hong Kong, Germany, France, Finland, Denmark, Dubai, Czech Republic, China, Korea, Belgium, Australia, and Amsterdam.
Product codes and lots
The device can be tracked according to service code. The service code for the DR 7500 system is 8087.