Z-2536-2014Class I
4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit
Abbott Medical / initiated 2014-07-07 / Terminated
Smiths Medical ASD, Inc. began a recall of 4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit on . FDA classified it as Class I and gives the reason as "The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a suction catheter". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2536-2014
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- 4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit
- Product code
- BTR Tube, Tracheal (W/Wo Connector)
- Reason for recall
- The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a suction catheter.
- FDA root cause
- Process change control
- Action
- Smiths Medical sent an Urgent Medical Device Recall Notice dated July 11, 2014, to all affected customers. The letter informed customers of the recall and instructed them to inspect their inventory and segregate any unused affected products and to complete and return the Recall Confirmation Form, by Fax 1-800-621-2590 or email to [email protected]. Customers with questions were instructed to call Smiths Medical's Customer Service Department at 1-800-258-5361. For questions regarading this recall call 866-216-8806.
- Quantity in commerce
- 135,518 (110,246 nationwide)
- Distribution
- Worldwide Distribution - USA including AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to CA, CY, GB, IE, IT, JE, KW, MT, PL, RU, SA, ZA.
- Product codes and lots
- Lot Numbers: 2470635, 2453453, 2443370, 2438608, 2412131, 2394050, 2380339, 2380338, 2376475, 2361590, 2352067, 2322810, 2271334, 2260548, 2223901, 2208177, 2202236, 2191909, 2161059, 2077168, 2042400, 12050
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2536-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2537-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2538-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2539-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2540-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2541-2014 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.