510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2536-2014Class I

4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit

Abbott Medical / initiated 2014-07-07 / Terminated

Smiths Medical ASD, Inc. began a recall of 4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit on . FDA classified it as Class I and gives the reason as "The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a suction catheter". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2536-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
4001PS Neonatal/Paed.Intubation Kit 2.5mm Murphy, LDS 15mm Conn Pediatric Endotracheal Intubation Kit
Product code
BTR Tube, Tracheal (W/Wo Connector)
Reason for recall
The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a suction catheter.
FDA root cause
Process change control
Action
Smiths Medical sent an Urgent Medical Device Recall Notice dated July 11, 2014, to all affected customers. The letter informed customers of the recall and instructed them to inspect their inventory and segregate any unused affected products and to complete and return the Recall Confirmation Form, by Fax 1-800-621-2590 or email to [email protected]. Customers with questions were instructed to call Smiths Medical's Customer Service Department at 1-800-258-5361. For questions regarading this recall call 866-216-8806.
Quantity in commerce
135,518 (110,246 nationwide)
Distribution
Worldwide Distribution - USA including AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to CA, CY, GB, IE, IT, JE, KW, MT, PL, RU, SA, ZA.
Product codes and lots
Lot Numbers: 2470635, 2453453, 2443370, 2438608, 2412131, 2394050, 2380339, 2380338, 2376475, 2361590, 2352067, 2322810, 2271334, 2260548, 2223901, 2208177, 2202236, 2191909, 2161059, 2077168, 2042400, 12050

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2536-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2536-2014Class ITerminatedVoluntary: Firm initiated
Z-2537-2014Class ITerminatedVoluntary: Firm initiated
Z-2538-2014Class ITerminatedVoluntary: Firm initiated
Z-2539-2014Class ITerminatedVoluntary: Firm initiated
Z-2540-2014Class ITerminatedVoluntary: Firm initiated
Z-2541-2014Class ITerminatedVoluntary: Firm initiated