Z-2564-2014Class II

Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to...

FUJIFILM Healthcare Americas Corporation / initiated 2014-08-18 / Terminated / root cause: design

Hitachi Medical Systems America Inc began a recall of Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical... on . FDA classified it as Class II and gives the reason as "The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2564-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard.
Root cause tag
design
FDA root cause
Device Design
Action
Letters will be sent to customers. Hatachi Serivce will be sent to each site to exchange the Gradient Coil Assembly.
Quantity in commerce
39 systems
Distribution
Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, ID, KS, KY, MD, MS, NY, OH, and WY; countries of: Japan, Brazil, France, Germany, and Kyrgyzstan.
Product codes and lots
Product Codes: Y001, Y002, Y003, Y004, Y005, Y006, Y007, Y008, Y009, Y010, Y011, Y012, Y014, Y015, Y016, Y101, Y102, Y103, Y105, and Y951. To be updated as firm submits information.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2564-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2564-2014Class IITerminatedVoluntary: Firm initiated