Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to...
Hitachi Medical Systems America Inc began a recall of Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical... on . FDA classified it as Class II and gives the reason as "The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2564-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Letters will be sent to customers. Hatachi Serivce will be sent to each site to exchange the Gradient Coil Assembly.
- Quantity in commerce
- 39 systems
- Distribution
- Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, ID, KS, KY, MD, MS, NY, OH, and WY; countries of: Japan, Brazil, France, Germany, and Kyrgyzstan.
- Product codes and lots
- Product Codes: Y001, Y002, Y003, Y004, Y005, Y006, Y007, Y008, Y009, Y010, Y011, Y012, Y014, Y015, Y016, Y101, Y102, Y103, Y105, and Y951. To be updated as firm submits information.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2564-2014 | Class II | Terminated | Voluntary: Firm initiated |