Z-2111-2014Class II

Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication...

Siemens Medical Solutions USA, Inc. / initiated 2014-06-30 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read... on . FDA classified it as Class II and gives the reason as "Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2111-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo Imaging XS also supports storage and archiving of Structured DICOM reports. Syngo Imaging XS optionally uses a variety of advanced post processing applications.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge enhancement and Histograms. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a Customer letter dated June 30th, to all affected customers to inform them of a new software version VA70B_0114 that will be introduced to all syngo Imaging XS customers with version VA70A or higher to solve a potential issue regarding miscalculations done for the grey scale values in the functions "Region of Interest", "Pixel Lens", "Edge Enhancement", and "Histograms". For questions regarding this recall call 610-219-6300.
Quantity in commerce
141
Distribution
Nationwide Distribution
Product codes and lots
model number: 10496279

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2111-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2111-2014Class IITerminatedVoluntary: Firm initiated