Z-2292-2014Class II

Siemens Luminos dRF, Ysio or Uroskop Omnia systems. The Axiom Luminos dRF is intended to be used as a...

Siemens Medical Solutions USA, Inc. / initiated 2014-07-07 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Luminos dRF, Ysio or Uroskop Omnia systems. The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and... on . FDA classified it as Class II and gives the reason as "It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2292-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Luminos dRF, Ysio or Uroskop Omnia systems. The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation exposure than required. This may also result in an overexposed image that is not of diagnostic quality. As a result the taken examination needs to be repeated. Thi
Root cause tag
design
FDA root cause
Device Design
Action
Siemens had informed all affected customers about this issue via Customer Safety Advisory Notice XP023/13/S (dated June 3, 2013) reported under C&R # 2240869-05/16/13-0013-C. A corresponding software update to fix the issue was originally planned to be rolled out by May 2014. The originally scheduled roll out date could not be met and to ensure that this new software version works accurately Siemens incorporated a comprehensive monitoring phase. Siemens has now scheduled the roll out for the software version VB10J to be completed by end of October 2014. A letter, dated July 7, 2014, with important safety information was sent to end users to notify them that the new software version VB10J will be introduced to solve the potential malfunction of the generator.
Quantity in commerce
192
Distribution
Nationwide Distribution.
Product codes and lots
model numbers 10094200, 10281163, 10281013, 10094910 with multiple serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2292-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2292-2014Class IITerminatedVoluntary: Firm initiated