Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2199-2014Class II

Prisma M100 PRE pump infusion set. Product number 107791. Each set is packaged in a plastic pouch, 4 pouches...

Gambro Renal Products / initiated 2014-07-18 / Terminated / root cause: design

Gambro Renal Products, Incorporated began a recall of Prisma M100 PRE pump infusion set. Product number 107791. Each set is packaged in a plastic pouch, 4 pouches per carton box on . FDA classified it as Class II and gives the reason as "Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2199-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gambro Renal Products
Product
Prisma M100 PRE pump infusion set. Product number 107791. Each set is packaged in a plastic pouch, 4 pouches per carton box
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Root cause tag
design
FDA root cause
Device Design
Action
Gambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Quantity in commerce
401,451
Distribution
Worldwide Distribution.
Product codes and lots
Prisma M100 PRE pump infusion set. Product number 107791. Batch number 12A2305 to 13D0902. Expiry 01/2014 to 04/2015

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2199-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2193-2014Class IITerminatedVoluntary: Firm initiated
Z-2194-2014Class IITerminatedVoluntary: Firm initiated
Z-2195-2014Class IITerminatedVoluntary: Firm initiated
Z-2196-2014Class IITerminatedVoluntary: Firm initiated
Z-2197-2014Class IITerminatedVoluntary: Firm initiated
Z-2198-2014Class IITerminatedVoluntary: Firm initiated
Z-2199-2014Class IITerminatedVoluntary: Firm initiated
Z-2200-2014Class IITerminatedVoluntary: Firm initiated
Z-2201-2014Class IITerminatedVoluntary: Firm initiated
Z-2202-2014Class IITerminatedVoluntary: Firm initiated
Z-2203-2014Class IITerminatedVoluntary: Firm initiated
Z-2204-2014Class IITerminatedVoluntary: Firm initiated
Z-2205-2014Class IITerminatedVoluntary: Firm initiated
Z-2206-2014Class IITerminatedVoluntary: Firm initiated
Z-2207-2014Class IITerminatedVoluntary: Firm initiated