Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2201-2014Class II

Prismaflex M60 set. Product number 106696. Each set is packaged in a plastic pouch, 4 pouches per carton box

Gambro Renal Products / initiated 2014-07-18 / Terminated / root cause: design

Gambro Renal Products, Incorporated began a recall of Prismaflex M60 set. Product number 106696. Each set is packaged in a plastic pouch, 4 pouches per carton box on . FDA classified it as Class II and gives the reason as "Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2201-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gambro Renal Products
Product
Prismaflex M60 set. Product number 106696. Each set is packaged in a plastic pouch, 4 pouches per carton box
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Root cause tag
design
FDA root cause
Device Design
Action
Gambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Quantity in commerce
401,451
Distribution
Worldwide Distribution.
Product codes and lots
Prismaflex M60 set. Product number 106696. Batch number 12A0605 to 14A2011. Expiry 01/2014 to 01/2016

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2201-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2193-2014Class IITerminatedVoluntary: Firm initiated
Z-2194-2014Class IITerminatedVoluntary: Firm initiated
Z-2195-2014Class IITerminatedVoluntary: Firm initiated
Z-2196-2014Class IITerminatedVoluntary: Firm initiated
Z-2197-2014Class IITerminatedVoluntary: Firm initiated
Z-2198-2014Class IITerminatedVoluntary: Firm initiated
Z-2199-2014Class IITerminatedVoluntary: Firm initiated
Z-2200-2014Class IITerminatedVoluntary: Firm initiated
Z-2201-2014Class IITerminatedVoluntary: Firm initiated
Z-2202-2014Class IITerminatedVoluntary: Firm initiated
Z-2203-2014Class IITerminatedVoluntary: Firm initiated
Z-2204-2014Class IITerminatedVoluntary: Firm initiated
Z-2205-2014Class IITerminatedVoluntary: Firm initiated
Z-2206-2014Class IITerminatedVoluntary: Firm initiated
Z-2207-2014Class IITerminatedVoluntary: Firm initiated