Z-2723-2014Class II

Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must...

Instrumed International Inc dba Avalign German Specialty Instruments / initiated 2014-05-30 / Terminated / root cause: labeling and UDI

Instrumed International, Inc. began a recall of Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to... on . FDA classified it as Class II and gives the reason as "Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2723-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Instrumed International Inc dba Avalign German Specialty Instruments
Product
Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Instrumed sent an URGENT: MEDICAL DEVICE CORRECTION letters dated May 30, 2014 to all consignees. The letters advised the consignees of the new warning statements on the product Instructions For Use and an instruction to complete and return the attached Response Request Form to Instrumed International. Consignees were further instructed to extend the recall to the retail level. For questions contact Instrumed Quality Manager at 847-908-6119.
Quantity in commerce
156 instruments
Distribution
Distribution US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
Product codes and lots
1) Instrumed Item No.: 250-08680; Boss Instruments Catalog Code: 35-8015; Lot Number: 030411, 100110, 110210;  2) Instrumed Item No.: 250-08700; Boss Instruments Catalog Code: 35-8016; Lot Number: 030111;  3) Instrumed Item No.: 250-08700; Boss Instruments Catalog Code: 35-8016; Lot Number: 30111;  4) Instrumed Item No.: 250-08760; Boss Instruments Catalog Code: 35-8020; Lot Number: 020411, 070111, 100412, 110412;  5) Instrumed Item No.: 250-08770; Boss Instruments Catalog Code: 35-8020L; Lot Number: 020112, 110211, 120310;  6) Instrumed Item No.: 250-08800; Boss Instruments Catalog Code: 35-8025; Lot Number: 060110, 100112;  7) Instrumed Item No.: 250-08820; Boss Instruments Catalog Code: 35-8026 ; Lot Number: 032811;  8) Instrumed Item No.: 250-08860; Boss Instruments Catalog Code: 35-8030; Lot Number: 050413, 070412, 110510, 090110;  9) Instrumed Item No.: 250-08880; Boss Instruments Catalog Code: 35-8030L; Lot Number: 090110;  10) Instrumed Item No.: 250-08890; Boss Instruments Catalog Code: 35-8035; Lot Number: 100213;  11) Instrumed Item No.: 250-08930; Boss Instruments Catalog Code: 35-8040; Lot Number: 010211, 020213, 020811, 021711, 040212, 040311, 080212, 100310, 110113, 120413;  12) Instrumed Item No.: 250-08960; Boss Instruments Catalog Code: 35-8040L; Lot Number: 030413, 040412, 040513, 120113;  13) Instrumed Item No.: 250-08990; Boss Instruments Catalog Code: 35-8042; Lot Number: 050312;  14) Instrumed Item No.: 250-09000; Boss Instruments Catalog Code: 35-8042L; Lot Number: 100406;  15) Instrumed Item No.: 250-09010; Boss Instruments Catalog Code: 35-8045; Lot Number: 010311, 020311, 090312, 100212, 120112, 120213

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2723-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2721-2014Class IITerminatedVoluntary: Firm initiated
Z-2722-2014Class IITerminatedVoluntary: Firm initiated
Z-2723-2014Class IITerminatedVoluntary: Firm initiated
Z-2724-2014Class IITerminatedVoluntary: Firm initiated
Z-2725-2014Class IITerminatedVoluntary: Firm initiated
Z-2726-2014Class IITerminatedVoluntary: Firm initiated
Z-2727-2014Class IITerminatedVoluntary: Firm initiated
Z-2728-2014Class IITerminatedVoluntary: Firm initiated
Z-2729-2014Class IITerminatedVoluntary: Firm initiated
Z-2730-2014Class IITerminatedVoluntary: Firm initiated