Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2562-2014Class II

Affinity Four Birthing Bed

Hill-Rom Inc. / initiated 2014-06-26 / Terminated / root cause: manufacturing process

Hill-Rom, Inc. began a recall of Affinity Four Birthing Bed, Product Usage: Product is intended to be used as birthing beds for women of child bearing age in an LDR (Labor, Delivery, Recovery)... on . FDA classified it as Class II and gives the reason as "The attachment latch mechanism can become bent depending on user handling. Latch mechanism bending could lead to improper engagement of the foot section and this could result in injury to the user due to a fall". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2562-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
Affinity Four Birthing Bed, Product Usage: Product is intended to be used as birthing beds for women of child bearing age in an LDR (Labor, Delivery, Recovery) or LDRP (Labor, Delivery, Recovery, Postpartum) setting within the acute care Labor and Delivery market.
Product code
KNC Table, Obstetric (And Accessories)
Reason for recall
The attachment latch mechanism can become bent depending on user handling. Latch mechanism bending could lead to improper engagement of the foot section and this could result in injury to the user due to a fall.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Hill-rom issued an Urgent Field Safety Notice letter dated June 26, 2014 to affected customers. The Notification identified the affected product issue and instructions for product inspection. The letter states that the firm is working on making changes to the Lift-Off foot section that will mitigate risk even if damage was to occur. If you have any questions concerning the notice or requested procedure, please contact Hill-Rom Technical Support at 800-445-3720.
Quantity in commerce
13,000 units
Distribution
Worldwide Distribution - US Nationwide and the following countries: Argentina , Aruba, Australia, Austria, Bahamas, Bahrain, Belgium, Belize, Bermuda, Brazil, Brunei, Bulgaria, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Equatorial Guinea, Finland, France, Germany, Greece, Guam, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Malaysia, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turks and Caicos Islands, United Arab Emirates, United Kingdom, US Virgin Islands, Venezuela and Vietnam.
Product codes and lots
Model #'s: P3700B, P3700C, P3700D and P3700E

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2562-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2562-2014Class IITerminatedVoluntary: Firm initiated