Z-2235-2014Class II

Vital Sync RMS 1.0 when loaded with v2.4 software, packaged on a CD or other electronically controlled...

Medtronic, Inc. / initiated 2014-07-17 / Terminated / root cause: software

Nellcor Puritan Bennett Inc. (dba Covidien LP) began a recall of Vital Sync RMS 1.0 when loaded with v2.4 software, packaged on a CD or other electronically controlled device. CDs are packaged in clam shell jeweled cases... on . FDA classified it as Class II and gives the reason as "Covidien is conducting a field correction of the Vital Sync Virtual Patient Monitoring Platform and Informatics Manager due to a mismatch between the alarm priority on the medical devices and the Vital Sync system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2235-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Vital Sync RMS 1.0 when loaded with v2.4 software, packaged on a CD or other electronically controlled device. CDs are packaged in clam shell jeweled cases within a carton box. The Vital Sync Virtual Patient Monitoring Platform is intended to be used by healthcare professionals for the following purposes: 1) To remotely consult regarding a patients status; and 2) To remotely review near real-time patient data, waveforms and alarms in order to utilize this information to aid in clinical decisions and deliver patient care in a timely manner.
Product code
MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Reason for recall
Covidien is conducting a field correction of the Vital Sync Virtual Patient Monitoring Platform and Informatics Manager due to a mismatch between the alarm priority on the medical devices and the Vital Sync system.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Covidien, sent an "URGENT FIELD CORRECTIVE ACTION NOTICE" letter dated 7/16/14 to its customers on 7/17/14 via FedEx. The letter informs users that a representative will be in contact to schedule a software update.The letter also described the product, problem and actions to be taken. The customers were instructed to disregard the audible and visual alarm priority indicators and only use the alarm description text on the patient tile and on the patient detail page for alarm priority information, until the mismatch is corrected. The Vital Sync" Virtual Patient Monitoring Platform and Informatics Manager or Vital Sync" Remote Monitoring System may still be used until the software update is available. If you have questions about this letter or to report any issues with the Vital Sync" Virtual Patient Monitoring Platform and Informatics Manager or Vital Sync" Remote Monitoring System, contact our Technical Support Department at (800)635-5267, option 3, then option 1.
Quantity in commerce
9
Distribution
US distribution to states of: LA, IL, MS, CO, IN, GA, CA and TX.
Product codes and lots
Model # 10116080 ; Lot # PO643063

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2235-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2234-2014Class IITerminatedVoluntary: Firm initiated
Z-2235-2014Class IITerminatedVoluntary: Firm initiated