EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist...
Intuitive Surgical, Inc. began a recall of EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the... on . FDA classified it as Class II and gives the reason as "Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2171-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the da Vinci Si Surgical System (Model IS3000) for resection, transection and/or creation of anastomoses in General, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for the enzymatic cleaner used throughout the cleaning process was modified from pH-neutral to pH-neutral to mild alkaline (mild alkaline is defined as pHd11). Mild Alkaline enzymatic
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Intuitive Surgical representatives hand delivered corrected manuals to the two affected sites..
- Quantity in commerce
- 2
- Distribution
- Distribution to Florida and South Carolina
- Product codes and lots
- Model number 410298-05 and 410298-06
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2171-2014 | Class II | Terminated | Voluntary: Firm initiated |