Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS...
Zimmer, Inc. began a recall of Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM on . FDA classified it as Class II and gives the reason as "Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2546-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Knee Instrument Fixation Screws: 48MM HEADED SCREW 48MM HEADLESS SCREW MIS HEADED SCREW, 48MM MIS HEADLESS SCREW, 48MM
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice.
- Quantity in commerce
- 972,125 units
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Item Numbers 00579104100, 00579104200, 00598304048, 00598304148 Exp Date Prior to 2023-11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2546-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2547-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2548-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2549-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2550-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2551-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2552-2014 | Class II | Terminated | Voluntary: Firm initiated |