Z-2236-2014Class II

Verigene BC-GN Test Kit (Catalog number 20-005-021) includes: 1) 20 Verigene BC-GN Test Cartridges. Each Test...

Luminex Corporation / initiated 2014-07-09 / Terminated

Nanosphere, Inc. began a recall of Verigene BC-GN Test Kit (Catalog number 20-005-021) includes: 1) 20 Verigene BC-GN Test Cartridges. Each Test Cartridge comes preloaded with all required... on . FDA classified it as Class II and gives the reason as "Nanosphere Inc. has recently determined through four customer complaints that a small number of Verigene Gram-Negative Blood Culture Nucleic Acid Test (BC-GN) Extraction Trays in Lot 042914021B have low levels or are...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2236-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Luminex Corporation
Product
Verigene BC-GN Test Kit (Catalog number 20-005-021) includes: 1) 20 Verigene BC-GN Test Cartridges. Each Test Cartridge comes preloaded with all required reaction solutions, including wash solutions, oligonucleotide probe solution and signal amplification solutions required to generate a test result. BC-GN is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, is not used to monitor these infections.
Product code
PEN Gram-Negative Bacteria And Associated Resistance Markers
Reason for recall
Nanosphere Inc. has recently determined through four customer complaints that a small number of Verigene Gram-Negative Blood Culture Nucleic Acid Test (BC-GN) Extraction Trays in Lot 042914021B have low levels or are missing Magnetic Microparticles (MMPs) required for DNA Extraction.
FDA root cause
Under Investigation by firm
Action
Nanosphere sent an Urgent Product Recall letter dated July 9, 2014 to all affected customers. All customers were initially notified of the recall by telephone on July 9, 2014. The customers were instructed to quarantine and return all affected product to Nanosphere. The letter included instructions: 1) on how to return the recalled products, and 2) to review any test results that were generated while using the recalled products to ensure that there has been no negative impact on patient care. Customers with questions were instructed to contact Nanosphere at 888-837-4436, option 2 or via e-mail at [email protected].
Quantity in commerce
1045 extraction trays (840 affected by recall)
Distribution
Nationwide Distribution including CA, CO, IA, MT, OH, OK, OR, TX, and UT
Product codes and lots
VERIGENE BC-GN TEST KIT (Catalog Number: 20-005-021). TEST CARTRIDGES: Part Number: 20-006-021, Lot Number: 042914021B, Expiration Date: 10/26/14

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2236-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2236-2014Class IITerminatedVoluntary: Firm initiated