Medtronic Ascenda Intrathecal Catheter models 8780 and 8781, and Ascenda Intrathecal Catheter Pump Segment...
Medtronic Neuromodulation began a recall of Medtronic Ascenda Intrathecal Catheter models 8780 and 8781, and Ascenda Intrathecal Catheter Pump Segment Revision Kit, model 8784. The Ascenda Intrathecal... on . FDA classified it as Class II and gives the reason as "Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2172-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic Ascenda Intrathecal Catheter models 8780 and 8781, and Ascenda Intrathecal Catheter Pump Segment Revision Kit, model 8784. The Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Ascenda Intrathecal Catheter.
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Medtronic sent an Urgent Medical Device Removal letter dated July 11, 2014, to all affected customers. The letter described the issue, identified affected product, provided patient management recommendations for patients implanted with an affected device, asked for Risk Management to quaratine unused inventory within the hospital, and to contact Medtronic Customer Service at 1-888-638-7627 to facilitate return of devices.
- Quantity in commerce
- 1200 (551 US, 649 OUS)
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of AL,AR,AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, Ml, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WV., and the countries of Austria, Belgium, Cyprus, Denmark, France, Germany, Iraq, Italy, Lebanon, Netherlands, Panama, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Kingdom and Uruguay.
- Product codes and lots
- Serial Numbers are in the following format: "N" + 9 digits (e.g. N12345001. Product beginning with the following letter/number combinations are affected: N464184, N464200, N464206, N464254, N464257, N464258, N464262, N464263, N464264, N464271, N464693, N464703, N464708, N464725, N464727, N464730, N464734, N464736, N464738, N464742, N464746, N464750, N464752, N464841, N464924, N464930, N464995, N464995, N465020, N465038, N465063, N465065, N465073, N465082, N465088, N465091, N465094, N465111, N465125, N465129, N465131, N465136, N465138, N465144, N465145, N465146, N465148, N465150, N466065, N466067, N466067, N466068, N466069, N466071, N466072, N466075, N466134, N466167, N466177, N466196, N466201, N466240, N466244, N466259, N466289, N466290, N466291, N466306, N466368, N466390, N466392, N466394, N466398, N466399, N466631, N466639, N466645, N466678, N466678, N466754, N466758, N466761, N466886, N466897, N466901, N466902, N466911, N466914, N467103, N467109, N467113, N467114, N467116, N467126, N467138, N467171, N467180, N467183, N467192, N467193, N467194, N467195, N467197, N467198, N467203, N467204, N467206, N467209, N467351, N467351, N467352, N467355, N467356, N467357, N467395, N467399, N467401, N467402, N467407, N467444, N467451, N467453, N467454, N467563, N467568, N467570, N467601, N467606, N467608, N467609, N467611, N467612, N467614, N467618, N467621, N467626, N468288, N468343, N468401, N468404, N468406, N468406, N468448, N468494, N468500, N468507, N468510, N468511, N468514, N468516, N468519, N468522, N468574, N468580, N468621, N468622, N468624, N468626, N468641, N469182, N469262, N469332, N469360, N469360, N469388, N469420, N469422, N469428, N469431, N469433, N469434, N469436, N469438, N469440, N469446, N469447, N469449, N469452, N469460, N469465, N469466, N469469, N469474, N469478, N469507, N469511, N469530, N469533, N469549, N469579, N469583, N469586, N469588, N470163, N470197, N470234, N470303, N470307, N470401, N470408, N470414, N470420, N470424, N470453, N470459, N470919, N470923, N470929, N470931, N470931, N470934, N470938, N470940, N470941, N470951, N470953, N470954, N471442, N471443, N471446, N471447, N471448, N471449, N471487, N471488, N471490, N471490, N471491, N471764, N471791, N471840, N471853, N471856, N471860, N471863, N471864, N471865, N471866, N471870, N471871, N471875, N471880, N472310, N472326, N472333, N472336, N472342, N472345, N472347, N472857, N472861, N472866, N472868, N472873, N472875, N472876, N472878, N472882, N472885, N472886, N472890, N472894, N472900, N472905, N473140, N473144, N473148, N473154, N473156, N473176, N473178, N473180, N473202, N473207, N473211, N473221, N473222, N473223, N473228, N473230, N473311, N473311, N473323, N473377, N473401, N473426, N473438, N473492, N473493, N473989, N473996, N473999, N474001, N474005, N474006, N474007, N474010, N474018, N474019, N474023, N474027, N474037, N474038, N474047, N474047, N474421, N474423, N474427, N474431, N474433.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2172-2014 | Class II | Terminated | Voluntary: Firm initiated |