510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2614-2014Class II

***Baxter***BURETROL Solution Set***105" (2.7 m) 150 mL Burette***Drip Chamber Filter Valve***3 Luer...

Baxter Healthcare Corporation / initiated 2014-07-14 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp began a recall of ***Baxter***BURETROL Solution Set***105" (2.7 m) 150 mL Burette***Drip Chamber Filter Valve***3 Luer Activated Valves***Male Luer Lock Adapter***with... on . FDA classified it as Class II and gives the reason as "Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2614-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
***Baxter***BURETROL Solution Set***105" (2.7 m) 150 mL Burette***Drip Chamber Filter Valve***3 Luer Activated Valves***Male Luer Lock Adapter***with Retractable Collar***Fluid path is Sterile, Nonpyrogenic. ****Indications for Use: For the administration of fluids from a container into the patient's vascular system through a vascular access device.***
Product code
FPA Set, Administration, Intravascular
Reason for recall
Baxter Healthcare Corporation is voluntarily issuing a recall for specific lots of Buretrol Solution set product codes due to complaints for separation between the burette chamber and the drip chamber. Affected product codes are: 2C7564, 2C8864 and 2H8864.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Action to be taken if product was purchased directly from Baxter: Baxter is requesting that you take the following actions: 1. Locate and remove all affected product from your facility (the product code and lot number can be found on the individual product package or shipping carton). 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. The Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time. Please have your Baxter eight-digit ship-to account number ready when calling. 3. Complete the enclosed customer reply form, and return it to Baxter by either fax Or scanned e-mail. 4. If you are a dealer, wholesaler, or distributor/reseller that distributed any product to other facilities, please conduct a recall with your end-user customers in accordance with your customary procedures. Baxter distributed this product to customers between September 27, 2013, and April 25, 2014. Action to be taken if the products were purshased from a distributor or re-seller. 1. Locate and remove all affected product from your facility (the product code and lot number can be found on the individual product package or shipping carton). 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. The Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time. Please have your ship-to account number ready when calling. 3. Follow your supplier's reply and recall process. Please do not return the customer reply form to Baxter.
Quantity in commerce
107232
Distribution
Nationwide Distribution.
Product codes and lots
2C8864- Buretrol Solution Set with 150mL CLEARLINK Burette, lots:  DR13E08015 DR13E08023 DR13E09013 DR13I26027 DR13I27017 DR13I27025 DR13I28015 DR13I28023 DR13J16027 DR13J17025 DR13J28022 DR13J29020 DR13K01019 DR13K20027 DR13K21025 DR13L19019 DR13L19027 DR14A11017 DR14A14011 DR14A24010 DR14A24036 DR14A31049 DR14B07054 DR14B22012 DR14B28050 DR14C01030 DR14C06013 DR14C06039

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2614-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2613-2014Class IITerminatedVoluntary: Firm initiated
Z-2614-2014Class IITerminatedVoluntary: Firm initiated
Z-2615-2014Class IITerminatedVoluntary: Firm initiated