Z-0020-2015Class II

Proteus 235 Proton Therapy System for cancer treatment

Ion Beam Applications S.A. / initiated 2014-03-27 / Terminated / root cause: design

Ion Beam Applications S.A. began a recall of Proteus 235 Proton Therapy System for cancer treatment on . FDA classified it as Class II and gives the reason as "While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0020-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ion Beam Applications S.A.
Product
Proteus 235 Proton Therapy System for cancer treatment.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a hot spot on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.
Root cause tag
design
FDA root cause
Device Design
Action
A User Notice was sent to direct accounts in/about March/April 2014 to notify them of the issue and provide instructions to avoid the identified potential risk to patients.
Quantity in commerce
8
Distribution
Worldwide Distribution-USA (nationwide) including the states of PA, NJ, IL, WA, and TN, and the countries of Germany, Italy, and Czech Republic.
Product codes and lots
PAT.107, PAT.109, PAT.112, PAT.113, PAT.114, PAT.115, PAT.116, PAT.119.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0020-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0020-2015Class IITerminatedVoluntary: Firm initiated