Z-2553-2014Class II

Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety...

Siemens Healthcare Diagnostics Inc. / initiated 2014-07-16 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of... on . FDA classified it as Class II and gives the reason as "The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2553-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens notified end users of a potential safety issue with the Reagent Lid hinge on the Dimension EXL instrument via an Urgent Medical Device Correction letter, dated July 2014. The letter indicated Customer Service Engineers will inspect and adjust the tension on the reagent lid hinges during every visit and replace hinges which no longer properly adjust. Users were urged to use caution when the Reagent Lid is open and notify Siemens the Reagent Lid is shifting down.
Quantity in commerce
2757
Distribution
Nationwide and internationally, including Canada and Mexico
Product codes and lots
All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2553-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2553-2014Class IITerminatedVoluntary: Firm initiated