510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2226-2014Class II

Double Offset Rasp Handle, Left Hand A/S Rasp Handle (Left)

Zimmer GmbH / initiated 2014-07-23 / Terminated / root cause: design

Zimmer, Inc. began a recall of Double Offset Rasp Handle, Left Hand A/S Rasp Handle (Left) on . FDA classified it as Class II and gives the reason as "The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2226-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Double Offset Rasp Handle, Left Hand A/S Rasp Handle (Left)
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring.
Root cause tag
design
FDA root cause
Device Design
Action
On July 23, 2014, Zimmer issued recall notification to their consignees. Distributors were notified via electronic mail and courier. Recall notification included a description of the issue, affected product, possible risks to patient, and instructions for responding to the recall. Questions or concerns regarding the recall are directed to Zimmer customer call center @ 1-800-348-2759 M-F, 8am-8pm ET.
Quantity in commerce
412 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of AK, AL, AZ, CA, CO, FL, GA, IN, KS, KY, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, and VA, and the countries of Australia, France, Germany, Italy, Switzerland, and Canada.
Product codes and lots
Item Number 00780803521  Lot Number 56474891, 56474892, 56474893, 56474894

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2226-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2216-2014Class IITerminatedVoluntary: Firm initiated
Z-2217-2014Class IITerminatedVoluntary: Firm initiated
Z-2218-2014Class IITerminatedVoluntary: Firm initiated
Z-2219-2014Class IITerminatedVoluntary: Firm initiated
Z-2220-2014Class IITerminatedVoluntary: Firm initiated
Z-2221-2014Class IITerminatedVoluntary: Firm initiated
Z-2222-2014Class IITerminatedVoluntary: Firm initiated
Z-2223-2014Class IITerminatedVoluntary: Firm initiated
Z-2224-2014Class IITerminatedVoluntary: Firm initiated
Z-2225-2014Class IITerminatedVoluntary: Firm initiated
Z-2226-2014Class IITerminatedVoluntary: Firm initiated
Z-2227-2014Class IITerminatedVoluntary: Firm initiated
Z-2228-2014Class IITerminatedVoluntary: Firm initiated
Z-2229-2014Class IITerminatedVoluntary: Firm initiated
Z-2230-2014Class IITerminatedVoluntary: Firm initiated
Z-2231-2014Class IITerminatedVoluntary: Firm initiated