VITROS Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS CA 19-9 Reagent Packs...
ORTHO-CLINICAL DIAGNOSTICS began a recall of VITROS Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS CA 19-9 Reagent Packs are used on VITROS ECi/ECiQ Immunodiagnostic... on . FDA classified it as Class II and gives the reason as "Customers may receive positively biased results when using VITROS Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2183-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS CA 19-9 Reagent Packs are used on VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic Systems and VITROS 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.
- Product code
- NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer
- Reason for recall
- Customers may receive positively biased results when using VITROS Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Ortho Clinical Diagnostics sent an Urgent Product Correction Notification July 2014 to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to Immediately discontinue using and discard Lots 1320 of VITROS CA 19-9 Calibrators and Reagent Packs. Review previously reported VITROS 25-OH Vitamin D Total results using Lot 1320. Review quality control results used to verify calibration events. Results lower than the target mean may indicate potentially biased assay results. Discuss any concerns regarding previously reported results with the Laboratory Medical Director or with the requesting physician. Complete and return the attached Confirmation of Receipt form. OCD will: credit their account for all VITROS CA 19-9 Calibrators and Reagent Packs that were discarded or ship repalcment product uspon request. Customers were instructed to forward the information in this notification if they have distributed this product outside of their facility. Customers with questions were instructed to contact their Customer Technical Service representative or contact the Customer Technical Service representative at 1-800-421-3311.
- Quantity in commerce
- 475
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to France and Japan
- Product codes and lots
- Lot 1320, Product code: 6800035 and 6800040
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2183-2014 | Class II | Terminated | Voluntary: Firm initiated |