510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2247-2014Class II

6.5 mm Cancellous Screw 16 mm Thread Length Sterile zimmer

Zimmer GmbH / initiated 2014-07-22 / Terminated / root cause: packaging

Zimmer, Inc. began a recall of 6.5 mm Cancellous Screw 16 mm Thread Length Sterile zimmer on . FDA classified it as Class II and gives the reason as "Zimmer is recalling various trauma screws that failed a leak test due to damage of the inner cavity only". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2247-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
6.5 mm Cancellous Screw 16 mm Thread Length Sterile zimmer
Product code
HWC Screw, Fixation, Bone
Reason for recall
Zimmer is recalling various trauma screws that failed a leak test due to damage of the inner cavity only.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
On 7/22/14, Zimmer sent recall notification to the consignees explaining the issue, identifying affected product, providing the health risks, and instructing the consignees what to do for the recall. Questions or concerns, please contact Zimmer at 1-877-946-2761.
Quantity in commerce
47,682 units total
Distribution
National Distribution: AK, AL, AZ, CA, CT, DC, FL, GA, US VIRGIN ISLANDS, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WI, WV, WY Foreign Distribution: Canada, Mexico, ARGENTINA, AUSTRALIA, BELGIUM,BRAZIL, CHINA, EL SALVADOR, ENGLAND, GERMANY, INDIA, ITALY, JAPAN, JORDAN, KOREA, NEW ZEALAND, SAUDI ARABIA, SINGAPORE, SPAIN, SWITZERLAND, TAIWAN, UAE
Product codes and lots
Item Number 47-4865-030, 035, 040, 045, 050, 055, 060, 065, 070, 075, 080, 085, 090, 100-01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2247-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2244-2014Class IITerminatedVoluntary: Firm initiated
Z-2245-2014Class IITerminatedVoluntary: Firm initiated
Z-2246-2014Class IITerminatedVoluntary: Firm initiated
Z-2247-2014Class IITerminatedVoluntary: Firm initiated
Z-2248-2014Class IITerminatedVoluntary: Firm initiated
Z-2249-2014Class IITerminatedVoluntary: Firm initiated
Z-2250-2014Class IITerminatedVoluntary: Firm initiated
Z-2251-2014Class IITerminatedVoluntary: Firm initiated