Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2256-2014Class II

Tibila Insert UC, Balanced Knee System, REF163-3YYY

Ortho Development Corp. / initiated 2014-07-11 / Terminated / root cause: packaging

Ortho Development Corporation began a recall of Tibila Insert UC, Balanced Knee System, REF163-3YYY Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for... on . FDA classified it as Class II and gives the reason as "Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2256-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho Development Corp.
Product
Tibila Insert UC, Balanced Knee System, REF163-3YYY Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.
Quantity in commerce
13
Distribution
Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Product codes and lots
UC Tibial Insert Sz 3 10mm,Part 163-3310,Lot A136109,UC Tibial Insert Sz 4 8mm,Part 163-3408,Lot A135948,UC Tibial Insert Sz 4 14mm,Part 163-3414,Lot A136110,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136113,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136746,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2256-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2253-2014Class IITerminatedVoluntary: Firm initiated
Z-2254-2014Class IITerminatedVoluntary: Firm initiated
Z-2255-2014Class IITerminatedVoluntary: Firm initiated
Z-2256-2014Class IITerminatedVoluntary: Firm initiated
Z-2257-2014Class IITerminatedVoluntary: Firm initiated