510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2212-2014Class II

Device Brand Name: Distal Radius Fracture Repair System Device Classification Name: Plate, Fixation, Bone...

Zimmer GmbH / initiated 2014-06-23 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Device Brand Name: Distal Radius Fracture Repair System Device Classification Name: Plate, Fixation, Bone Device Common Name: Drill Guide DNP 3.3mm Labeling... on . FDA classified it as Class II and gives the reason as "The firm received complaints indicating that the drill guide becomes stuck in the jig. The drill guide can be placed in the jig about halfway before it becomes stuck". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2212-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Device Brand Name: Distal Radius Fracture Repair System Device Classification Name: Plate, Fixation, Bone Device Common Name: Drill Guide DNP 3.3mm Labeling: DRILL GUIDE DNP 3.3 MM NON-STERILE LOT NB0311 QTY 1 DePuy P.O Box 988 Warsaw IN 46581
Product code
HRS Plate, Fixation, Bone
Reason for recall
The firm received complaints indicating that the drill guide becomes stuck in the jig. The drill guide can be placed in the jig about halfway before it becomes stuck.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On June 23, 2014 Biomet issued recall notification to their consignees via letter. Notification included affected product issue and instructions for return were provided. Distributors who had affected hospitals in their regions were notified and provided with a copy of the notice that was sent to the hospital. Consignees are being contacted on an individual basis in an attempt to reconcile the product 100%. A response form was provided to all consignees in order for Biomet to track returns. Non-responders will be contacted on an individual basis.
Quantity in commerce
37
Distribution
The product was distributed domestically and to internationally. Domestically to CA, PA, IA, CO, NY, MD, NC,AL, AZ, IL, GA, VA. Netherlands, Argentina,
Product codes and lots
Catalog Number: DNPDG, Lot Number: NB0311 and NB0611

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2212-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2212-2014Class IITerminatedVoluntary: Firm initiated