Medtronic, Activa PC, Model 37601, Method of Sterilization: Ethylene Oxide, Single Use Only, Rx Only. The...
Medtronic Neuromodulation began a recall of Medtronic, Activa PC, Model 37601, Method of Sterilization: Ethylene Oxide, Single Use Only, Rx Only. The Activa PC neurostimulator is a dual-channel device... on . FDA classified it as Class II and gives the reason as "Medtronic is recalling six Activa PC (model 37601) Implantable Neurostimulators due to the potential for a damaged electrical component during manufacturing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2259-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic, Activa PC, Model 37601, Method of Sterilization: Ethylene Oxide, Single Use Only, Rx Only. The Activa PC neurostimulator is a dual-channel device capable of delivering bilateral stimulation. Activa PC contains a non-rechargeable battery and microelectronic circuitry to deliver a controlled electrical pulse to precisely targeted areas of the brain. The device is typically implanted subcutaneously near the clavicle, connected to an extension and leads, which are implanted in the brain.
- Product code
- MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Reason for recall
- Medtronic is recalling six Activa PC (model 37601) Implantable Neurostimulators due to the potential for a damaged electrical component during manufacturing.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm, Medtronic, notified their Consignees on 07/14/2014 via telephone of the recall. Medtronic representative used telephone script to convey the information. The script was directed to Risk Management or Inventory Management. The caller was to inform consignees of the problem and product being recalled. Advised consignees to quarantine the product and provided the Medtronic Device Removal Reply Form to the consignees via e-mail or Fax. The Reply Form included contact information which was to call 1-800-633-8766 in case they needed to contact a Medtronic representative. The completed form is to be faxed back to 1-800-897-3899 or e-mail a PDF to [email protected]. If you have any questions, call 763-526-1294.
- Quantity in commerce
- 6
- Distribution
- Distributed in the states of: MA, NC, OH, and TX.
- Product codes and lots
- Serial numbers: NKM724776H, NKM724782H, NKM724785H, NKM724790H, NKM724802H, NKM724843H.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2259-2014 | Class II | Terminated | Voluntary: Firm initiated |