Galt Valved Tearaway Introducer Set Galt Vti Rx Only Units 7 Sterile Eo
Galt Medical Corporation began a recall of GALT VALVED TEARAWAY INTRODUCER SET GALT VTI RX ONLY UNITS 7 STERILE EO Product Usage: Used in percutaneous procedures to introduce catheters and other... on . FDA classified it as Class II and gives the reason as "The adhesive used in the product could dislodge when objects are inserted through the valve which could pose an embolic risk to the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2588-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Galt Medical Corp
- Product
- GALT VALVED TEARAWAY INTRODUCER SET GALT VTI RX ONLY UNITS 7 STERILE EO Product Usage: Used in percutaneous procedures to introduce catheters and other intravascular devices into the peripheral vasculature.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- The adhesive used in the product could dislodge when objects are inserted through the valve which could pose an embolic risk to the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Galt Medical sent an Urgent Medical Device Recall letters to all affected customers. The letter identified the affected product, reason for recall and actions to be taken. Customers were asked to identify and segregate the recalled lot (s) that is in their possession, complete and return the enclosed Recall Reply form, and return the recalled product to Galt Medical Corp. For questions contact your local sales representative or Recall Coordinator at 214-778-1306.
- Quantity in commerce
- 11,155 units in total
- Distribution
- Worldwide Distribution - US (nationwide) Canada, Germany, the Netherlands, Ireland, Switzerland, France, Italy, and Greece.
- Product codes and lots
- Catalog Numbers: NT-104-06, INT-104-07, INT-104-08, INT-104-09, INT-104-10, INT-104-11, INT-104-12, INT-104-17, INT-104-18, INT-104-19, INT-104-20, INT-104-21, INT-104-22, INT-104-23, INT-108-06, INT-108-07, INT-108-09, INT-108-10, INT-108-11, INT-108-12, INT-108-17, INT-108-18, INT-108-19, INT-108-20, INT-108-21, INT-108-22, INT-108-22, VTI-002-06, VTI-002-07, VTI-002-08, VTI-002-09, VTI-002-09, VTI-002-10, VTI-002-11, VTI-002-12, VTI-002-17, VTI-002-18, VTI-002-19, VTI-002-20, VTI-002-21, VTI-002-22, VTI-002-23, VTI-003-06, VTI-003-07, VTI-003-08, VTI-003-09, VTI-003-10, VTI-003-11, VTI-003-12, VTI-003-17, VTI-003-18, VTI-003-19, VTI-003-20, VTI-003-21, VTI-003-22, and VTI-003-23.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2588-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2589-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY