Persona (TM) The Personalized Knee System CR Tibial Articular Surface Provisional Top Nonsterile TASP CR L...
Zimmer, Inc. began a recall of Persona (TM) The Personalized Knee System CR Tibial Articular Surface Provisional Top Nonsterile TASP CR L 3-9 CD TOP on . FDA classified it as Class II and gives the reason as "Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2298-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Persona (TM) The Personalized Knee System CR Tibial Articular Surface Provisional Top Nonsterile TASP CR L 3-9 CD TOP
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On February 10, 2014, Zimmer sent letter to surgeons who had filed complaints regarding the affected TASP fractures. The letter provided the surgeons with Zimmer's investigational findings of the issue. Surgeons were directed to send questions or any additional informtion to [email protected]. The Zimmer Weekly Wrap-Up, dated February 14, 2014, was issued electronically to sales force. They were notified to engage in mandatory training to improve the usage of the TASPs and prevent potential breakage. Zimmer has taken additional corrective measures by initiating another recall in July 2014 whereby they notify the distributors as well as all users of the enhanced surgical technique.
- Quantity in commerce
- 70,986 distribution events
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY, and the countries of Australia, Canada, Germany, India, Japan, Korea, and Singapore.
- Product codes and lots
- Item Numbers: 42517000410 Lot Numbers: 62024975 62043546 62144792 62152725 62165538 62187318 62337961 62372407 62386776 62397159 62405917 62456263 62499170 62551386 62568272 62572572 62565051 62663168 62691384 62707494 62717346
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2296-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2297-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2298-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2299-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2300-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2301-2014 | Class II | Terminated | Voluntary: Firm initiated |