510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2299-2014Class II

Persona (TM) The Personalized Knee System UC Tibial Articular Surface Provisional Top Nonsterile TASP UC L...

Zimmer GmbH / initiated 2013-06-10 / Terminated / root cause: design

Zimmer, Inc. began a recall of Persona (TM) The Personalized Knee System UC Tibial Articular Surface Provisional Top Nonsterile TASP UC L 3-7 CD TOP TASP UC L 4-11 EF TOP TASP UC L 7-12 GH... on . FDA classified it as Class II and gives the reason as "Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2299-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Persona (TM) The Personalized Knee System UC Tibial Articular Surface Provisional Top Nonsterile TASP UC L 3-7 CD TOP TASP UC L 4-11 EF TOP TASP UC L 7-12 GH TOP TASP UC L 9-12 J TOP
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.
Root cause tag
design
FDA root cause
Device Design
Action
On February 10, 2014, Zimmer sent letter to surgeons who had filed complaints regarding the affected TASP fractures. The letter provided the surgeons with Zimmer's investigational findings of the issue. Surgeons were directed to send questions or any additional informtion to [email protected]. The Zimmer Weekly Wrap-Up, dated February 14, 2014, was issued electronically to sales force. They were notified to engage in mandatory training to improve the usage of the TASPs and prevent potential breakage. Zimmer has taken additional corrective measures by initiating another recall in July 2014 whereby they notify the distributors as well as all users of the enhanced surgical technique.
Quantity in commerce
70,986 distribution events
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY, and the countries of Australia, Canada, Germany, India, Japan, Korea, and Singapore.
Product codes and lots
Item Numbers: 42517200410 42517200510 42517200610 42517200710  Lot Numbers:  62034570 62117742 62156315 62187322 62247350 62314721 62372409 62381836 62397160 62429190 62474906 62551390 62565048 62595853 62614781 62700416 62702297 62724349 62043548 62055681 62104368 62117743 62144794 62165545 62177246 62256282 62351288 62356244 62391228 62427419 62456268 62506003 62551338 62564355 62597762 62614786 62702292 62705808 62717332 62055682 62068376 62068378 62144795 62152727 62187323 62256283 62351289 62386419 62356275 62445931 62456266 62558584 62575757 62611087 62638672 62698749 62330130 62563854 62567418

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2299-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2296-2014Class IITerminatedVoluntary: Firm initiated
Z-2297-2014Class IITerminatedVoluntary: Firm initiated
Z-2298-2014Class IITerminatedVoluntary: Firm initiated
Z-2299-2014Class IITerminatedVoluntary: Firm initiated
Z-2300-2014Class IITerminatedVoluntary: Firm initiated
Z-2301-2014Class IITerminatedVoluntary: Firm initiated