Z-2561-2014Class II

Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A...

Siemens Healthcare Diagnostics Inc. / initiated 2013-08-13 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data management system that assists medical... on . FDA classified it as Class II and gives the reason as "May not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2561-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions in conjunction with multiple instruments, the laboratory information system (LIS) and Siemens StreamLAB Automation Solutions.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
May not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued an Urgent Medical Device Correction (UMDC) letter on August 13, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review this letter in its entirety and ensure that all operators understand the information presented. The issues described are specific to the features listed in the letter.. If the laboratory is not using a particular feature, then they are not affected by the issues described for that feature. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Quantity in commerce
1347
Distribution
US Nationwide Distribution
Product codes and lots
EasyLink Informatics System V4.0/V5 Siemens Material Number (SMN) 10444799, 10444800, 10487219

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2561-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2561-2014Class IITerminatedVoluntary: Firm initiated