Trifurcated Transpac IV Monitoring Kit w/Safeset Reservoir, 03 ml Squeeze Flush and 6 Needleless Valves, Item...
ICU Medical, Inc. began a recall of Trifurcated Transpac IV Monitoring Kit w/Safeset Reservoir, 03 ml Squeeze Flush and 6 Needleless Valves, Item No. 011-46106-22 Transpac Trifurcated monitoring... on . FDA classified it as Class II and gives the reason as "ICU has become aware of an issue with a component where the connection on an arterial monitoring kit may crack and cause a fluid leak". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2160-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Trifurcated Transpac IV Monitoring Kit w/Safeset Reservoir, 03 ml Squeeze Flush and 6 Needleless Valves, Item No. 011-46106-22 Transpac Trifurcated monitoring kit with Safeset Reservoir, squeeze flush and sampling ports. The monitoring kit is intended for measuring and montoring of fluid pressure, fluid infusion, and blood withdrawal.
- Product code
- DRS Transducer, Blood-Pressure, Extravascular
- Reason for recall
- ICU has become aware of an issue with a component where the connection on an arterial monitoring kit may crack and cause a fluid leak.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- ICU Medical, Inc. sent an email dated July 18, 2014, to their customers, The email inform the customer of the problem and the action to be taken. The email stated that the US Sales Representative would remove the recalled product on July 22, 2014. If product was further distribution customers should notify their customers. Customer was informed of the RGA to have the recalled product returned. For questions regarding this recall call 801-264-1732.
- Quantity in commerce
- 60 units
- Distribution
- Worldwide Distribution - USA including Oregon and Internationally to Australia.
- Product codes and lots
- Lot No. 2848741, 2859329, 2862831, 2869453
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2157-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2158-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2159-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2160-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2161-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2162-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2163-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2164-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2165-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
