Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2671-2014Class II

VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G...

Ortho-Clinical Diagnostics, Inc. / initiated 2014-08-06 / Terminated / root cause: packaging

ORTHO-CLINICAL DIAGNOSTICS began a recall of VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody to hepatitis C virus (anti-HCV) in human... on . FDA classified it as Class II and gives the reason as "VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2671-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody to hepatitis C virus (anti-HCV) in human serum and plasma. REF 680 2454, IVD -- Ortho-Clinical Diagnostics High Wycombe UK. Note that one sales unit has 5 Reagent Packs contained within it. Each Reagent Pack can test for 100 samples (patient, Quality Control, etc.).
Product code
MZO Assay, Enzyme Linked Immunosorbent, Hepatitis C Virus
Reason for recall
VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in unexpected negative test results being obtained for reactive sample fluids (quality control and patient samples).
Root cause tag
packaging
FDA root cause
Packaging process control
Action
All US consignees (users) who were shipped the affected lot VITROS HCV Reagent Pack were sent communications via FedEx and/or US Priority Mail (PO Boxes only) on 8/06/14, and advised of the issue and to immediately discontinue using all remaining inventory of VITROS Anti-HCV Reagent Packs Lot 9090. For additional information, please contact OCD's Customer Technical Service representatives at 1-800-421-3311.
Quantity in commerce
463 sales units
Distribution
NY, NJ, CT, MA, PA, VA, MD, FL and GA
Product codes and lots
Lot 9090 (exp. December 15, 2014)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2671-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2671-2014Class IITerminatedVoluntary: Firm initiated