FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of...
Elekta, Inc. began a recall of FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of treatment plan outputs. Also able to perform CT... on . FDA classified it as Class II and gives the reason as "When DICOM is exporting an arc plan (VMAT, Dynamic Conformal of 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2574-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of treatment plan outputs. Also able to perform CT simulations and forward their information to the RTP System for dose calculation.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- When DICOM is exporting an arc plan (VMAT, Dynamic Conformal of 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment plan.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Elekta, Inc. issued an Urgent Important Field Safety Notice letter (IFSN 382-01-MON-001) to all affected customers.. Confirmation of receipt is to be completed by October 31, 2014. The problem will be resolved in a patch to Focal 4.80. Development is underway with a release date of September 2014 with confirmation of installation of the patch for the fix no later than 6 months from release of the patch. An updated notice from the manufacturer is also being sent to affected customers (FCO: 382-01-MON-001 VID:2.0). For questions contact your local Elekta office.
- Quantity in commerce
- 554 units
- Distribution
- Worldwide Distribution - US (nationwide)in the states of AK, CA, FL, MI, MN, MO, NJ, NY, NC, OH, PA, TN, VA,WA and Puerto Rico, Algeria, Antigua & Barbuda, Australia, Austria, Bahrain, Belarus, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech republic, Ecuador, Egypt, Estonia, France, Germany, Greece, India, Iraq, Israel, Italy, Japan, Kazakhstan, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Pakistan Portugal, Romania, Russia, Saudi Arabia, Singapore, Solvenia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, and United Kingdom
- Product codes and lots
- Software versions 4.70.00 & 4.80.00
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2574-2014 | Class II | Terminated | Voluntary: Firm initiated |