Z-2577-2014Class II

PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size...

Zimmer GmbH / initiated 2014-08-07 / Terminated / root cause: design

Zimmer, Inc. began a recall of PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH... on . FDA classified it as Class II and gives the reason as "Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2577-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Left Size 3-9 CD Top; Left Size 3-11 EF Top; Right Size 7-12 GH Top; and Right Size 9-12 J Top.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.
Root cause tag
design
FDA root cause
Device Design
Action
On 8/7/2014, Zimmer notified all distributors via electronic mail. Distributors that have received affected inventory are also notified via courier. Hospital risk managers and surgeons are also notified via courier. Hospital risk managers and surgeons were provided with a notification identifying the issue and their responsibilities. These responsibilities include ensuring the affected personnel are aware of the contents of the notification. Distributors were sent a notification identifying the issue and to provide a copy of the notification sent to surgeons and risk managers. The distributors will also be instructed as needed to assist with identifying a comprehensive list of surgeons.
Quantity in commerce
77,737 total units
Distribution
Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of: AUSTRALIA, INDIA, JAPAN, KOREA, SINGAPORE, AUSTRIA, BELGIUM, FRANCE, GERMANY, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWITZERLAND, UNITED KINGDOM, and UNITED ARAB EMIRATES.
Product codes and lots
Part Numbers: 42517000410, 42517000510, 42517000610, 42517000710, 42527000410, 42527000510, 42527000610, and 42527000710. Lot numbers: 62024975, 62043546, 62144792, 62152725, 62165538, 62187318, 62337961, 62372407, 62386776, 62397159, 62405917, 62456263, 62499170, 62551386, 62565051, 62568272, 62572572, 62663168, 62691384, 62707494, 62717346, 62020920, 62144793, 62156313, 62165539, 62184228, 62353932, 62356176, 62385100, 62420242, 62468277, 62538412, 62564340, 62611088, 62612263, 62690056, 62702295, 62717337, 62736078, 62024976, 62054776, 62147232, 62177243, 62187319, 62187320, 62359057, 62364955, 62391222, 62414006, 62461233, 62558583, 62564345, 62608889, 62638657, 62646286, 62702749, 62724348, 62394279, 62443397, 62487490, 62492802, 62506001, 62563915, 62568271, 62704633, 62055686, 62144796, 62161537, 62162597, 62256286, 62351297, 62356578, 62385099, 62404282, 62424802, 62455897, 62538392, 62565037, 62595855, 62611083, 62663162, 62680121, 62686205, 62695286, 62724347, 62724744, 62019863, 62156317, 62165548, 62168386, 62184268, 62256288, 62344258, 62356586, 62391232, 62391233, 62455892, 62504877, 62551337, 62564335, 62597816, 62638685, 62646284, 62695878, 62036673, 62138318, 62149715, 62189326, 62189327, 62356593, 62377738, 62391236, 62420225, 62461232, 62564360, 62564361, 62595865, 62611086, 62680120, 62690951, 62710615, 62761109, 62394280, 62428338, 62444959, 62500931, and 62568278.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2577-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2575-2014Class IITerminatedVoluntary: Firm initiated
Z-2576-2014Class IITerminatedVoluntary: Firm initiated
Z-2577-2014Class IITerminatedVoluntary: Firm initiated
Z-2578-2014Class IITerminatedVoluntary: Firm initiated
Z-2579-2014Class IITerminatedVoluntary: Firm initiated
Z-2580-2014Class IITerminatedVoluntary: Firm initiated