Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2206-2014Class II

PrisMARS Kit (component of MARS Treatment Kit Type 1115/1 PrisMARS US), 1 MARS Treatment Kit per carton box...

Gambro Renal Products / initiated 2014-07-18 / Terminated / root cause: design

Gambro Renal Products, Incorporated began a recall of PrisMARS Kit (component of MARS Treatment Kit Type 1115/1 PrisMARS US), 1 MARS Treatment Kit per carton box, Product Number 800480, Batch on . FDA classified it as Class II and gives the reason as "Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2206-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gambro Renal Products
Product
PrisMARS Kit (component of MARS Treatment Kit Type 1115/1 PrisMARS US), 1 MARS Treatment Kit per carton box, Product Number 800480, Batch
Product code
FLD Apparatus, Hemoperfusion, Sorbent
Reason for recall
Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Root cause tag
design
FDA root cause
Device Design
Action
Gambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Quantity in commerce
401,451
Distribution
Worldwide Distribution.
Product codes and lots
Model Number MARS Treatment Kit Type 1115/1 PrisMARS US, Batch 0000017412 - 0000019827, Expiry Date 06/2014 to 05/2016

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2206-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2193-2014Class IITerminatedVoluntary: Firm initiated
Z-2194-2014Class IITerminatedVoluntary: Firm initiated
Z-2195-2014Class IITerminatedVoluntary: Firm initiated
Z-2196-2014Class IITerminatedVoluntary: Firm initiated
Z-2197-2014Class IITerminatedVoluntary: Firm initiated
Z-2198-2014Class IITerminatedVoluntary: Firm initiated
Z-2199-2014Class IITerminatedVoluntary: Firm initiated
Z-2200-2014Class IITerminatedVoluntary: Firm initiated
Z-2201-2014Class IITerminatedVoluntary: Firm initiated
Z-2202-2014Class IITerminatedVoluntary: Firm initiated
Z-2203-2014Class IITerminatedVoluntary: Firm initiated
Z-2204-2014Class IITerminatedVoluntary: Firm initiated
Z-2205-2014Class IITerminatedVoluntary: Firm initiated
Z-2206-2014Class IITerminatedVoluntary: Firm initiated
Z-2207-2014Class IITerminatedVoluntary: Firm initiated