Z-2573-2014Class II
Streamline CT Cervico-Thoracic System; Set Screw Ref 26-Setscrew Non-Sterile, Single Use, Rx Only
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) / initiated 2012-09-10 / Terminated / root cause: manufacturing process
PIONEER SURGICAL TECHNOLOGY, INC. began a recall of STREAMLINE CT CERVICO-THORACIC SYSTEM; SET SCREW REF 26-SETSCREW NON-STERILE, SINGLE USE, RX ONLY on . FDA classified it as Class II and gives the reason as "Some set screw thread profiles were found not to be within specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2573-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
- Product
- Streamline CT Cervico-Thoracic System; Set Screw Ref 26-Setscrew Non-Sterile, Single Use, Rx Only
- Product code
- KWP Appliance, Fixation, Spinal Interlaminal
- Reason for recall
- Some set screw thread profiles were found not to be within specifications.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recall took place at the distributor/sales representative level. Communication was done through phone and email notifications on 9/10/2012. All affected set screws were to be traded out through a scheduled process. This was coordinated through Pioneer Surgical's customer service. Product was scrapped and no reconditioning took place. All affected set screws were swapped out.
- Quantity in commerce
- 11 sets
- Distribution
- Distributed in the states of NC, NH, NY, CA, MI, IL, and FL.
- Product codes and lots
- Lot numbers: 104875, 104876, 104877, 104878, 104879, 104880, 104881 , 117030,117031 , 121947, 124260.