Z-2681-2014Class II

The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result...

Luminex Corporation / initiated 2013-10-10 / Terminated / root cause: manufacturing process

Nanosphere, Inc. began a recall of The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro... on . FDA classified it as Class II and gives the reason as "Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2681-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Luminex Corporation
Product
The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram negative bacteria as determined by gram stain.
Product code
PEN Gram-Negative Bacteria And Associated Resistance Markers
Reason for recall
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.
Quantity in commerce
400 units
Distribution
Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.
Product codes and lots
Product Number: 20-009-021; Lot Number and Expiration Date: Lot: 100113021C, Expiration Date: 03/30/14

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2681-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2681-2014Class IITerminatedVoluntary: Firm initiated
Z-2682-2014Class IITerminatedVoluntary: Firm initiated
Z-2683-2014Class IITerminatedVoluntary: Firm initiated
Z-2684-2014Class IITerminatedVoluntary: Firm initiated
Z-2685-2014Class IITerminatedVoluntary: Firm initiated