Z-2616-2014Class II

Streamline MIS Spinal Fixation System Fixed Rod Holder NON STERILE, Rx only The Streamline MIS Spinal...

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) / initiated 2013-06-14 / Terminated / root cause: design

PIONEER SURGICAL TECHNOLOGY, INC. began a recall of Streamline MIS Spinal Fixation System Fixed Rod Holder NON STERILE, Rx only The Streamline MIS Spinal Fixation System is intended for posterior, noncervical... on . FDA classified it as Class II and gives the reason as "The affected fixed rod holders did not allow the instrument to lock down on the spinal rod". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2616-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Product
Streamline MIS Spinal Fixation System Fixed Rod Holder NON STERILE, Rx only The Streamline MIS Spinal Fixation System is intended for posterior, noncervical pedicle fixation, T1-S2. Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion. The device is indicated for all the following indications: degenerative disc disease (ODD) (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; deformities or curvatures (i.e. scoliosis, kyphosis, and /or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Streamline MIS instrumentation, when used with the Streamline MIS Spinal Fixation System, is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
The affected fixed rod holders did not allow the instrument to lock down on the spinal rod.
Root cause tag
design
FDA root cause
Device Design
Action
Pioneer Surgical sent a Product Field Action letter on July 7, 2013, to all affected customers. The letter described the affected product and issue, and provided instructions for return of the recalled product. All 9 affected instruments were returned to Pioneer Surgical, reconditioned and sent back to the field. Customers with questions were instructed to call 906-226-4489. For questions regarding this recall call 906-226-4812.
Quantity in commerce
9
Distribution
Nationwide Distribution including TX, CA, IL, and FL.
Product codes and lots
Lot number: 148062, 148440

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2616-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2616-2014Class IITerminatedVoluntary: Firm initiated