510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2636-2014Class II

Volcano TrakBack II Disposable Pullback Device; Sterile and nonpyrogenic; Single Use. TrakBack II is used to...

Volcano Corporation / initiated 2014-08-13 / Terminated / root cause: packaging

Volcano Corporation began a recall of Volcano TrakBack II Disposable Pullback Device; Sterile and nonpyrogenic; Single Use. TrakBack II is used to move the catheter steadily and precisely, allowing... on . FDA classified it as Class II and gives the reason as "Internal testing had identified a potential breach to the sterile barrier that may impact the sterility of the product". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2636-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Volcano Corporation
Product
Volcano TrakBack II Disposable Pullback Device; Sterile and nonpyrogenic; Single Use. TrakBack II is used to move the catheter steadily and precisely, allowing uniform collection of imaging data.
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Internal testing had identified a potential breach to the sterile barrier that may impact the sterility of the product.
Root cause tag
packaging
FDA root cause
Packaging
Action
Volcano sent an Urgent Product Recall Notification letter dated August 13, 2014, to all customers who received the device in the last 25 months. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately cease use of the affected product and return unused devices to the Volcano distribution center. Customers were asked to prepare the attached inventory log and then contact Volcano Customer Service at 800-228-4728, option 4 to arrange for return of any affected product in their inventory. A credit would be issued for any returned product. Customers with questions were asked to call 800-228-4728. For questions regarding this recall call 978-439-3586.
Quantity in commerce
8251 total devices (3777 US)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Korea, Thailand, Canada, Australia, Chili, Mexico, Argentina.
Product codes and lots
Part number 91003: All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2636-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2636-2014Class IITerminatedVoluntary: Firm initiated