510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0061-2015Class II

Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor...

Depuy Orthopaedics Inc. / initiated 2014-08-27 / Terminated

DePuy Orthopaedics, Inc. began a recall of Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor and Handle Set is used to facilitate the insertion... on . FDA classified it as Class II and gives the reason as "The small extraction peg of the Charnley Pin Retractor and Handle set is breaking from the pin". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0061-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Charnley PN Retract & Handle, REF 962004000, Non-Sterile-DuPuy International Ltd. The Charnley Pin Retractor and Handle Set is used to facilitate the insertion and extraction of two pins into and out of bone in the incision area, thereby holding the incision soft tissue open during the surgical procedure.
Product code
MDM Instrument, Manual, Surgical, General Use
Reason for recall
The small extraction peg of the Charnley Pin Retractor and Handle set is breaking from the pin.
FDA root cause
Mixed-up of materials/components
Action
The firm, Depuy Synthes, sent an "URGENT INFORMATION - RECALL NOTICE" dated August 27, 2014 to their consignees. The recall notification included a description of the reason for recall, affected product, clinical implications, and instructions for responding to the formal recall notification. The customers were instructed to cease using the affected components immediately, quarantine all affected components in a manner that ensures the affected components are not used; return affected devices immediately to your DePuy; if any affected product has been forwarded to another facility, contact that facility immediately to communicate recall and arrange return; maintain awareness and copy of this notice, and complete and return the Reconciliation forms via fax to 574-371-4939 or emailed to [email protected]. For questions about the device recall information provided, please contact Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT) or email [email protected]. For product-related questions, please contact your local DePuy Orthopaedics, Inc. Sales Consultant. For clinical questions from hospitals/user facilities, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT).
Quantity in commerce
65 units
Distribution
Worldwide Distribution: US (Nationwide) including states of: AR, CA, CO, CT, IL, MA, NC, OH, PA, TN, TX, UT, VA, WA, WI; and Internationally to: AUSTRALIA, CHILE, UK, INDIA, IRELAND, JAPAN, NEW ZEALAND, RUSSIA, SINGAPORE, and SAUDI ARABIA.
Product codes and lots
Barcode: 10603295240532 CATALOG NO. : 962004000  Lot # OSA-149328 Lot # OSA-181610  Lot # OSA-161157

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0061-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0061-2015Class IITerminatedVoluntary: Firm initiated