Siemens Dimension Vista CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the...
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension Vista CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL)... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2720-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Dimension Vista CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea Nitrogen (BUN), and Uric Acid (UA) methods on the Dimension Vista System.
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista® CHEM 1 CAL, lot 3GM081. Investigation by Siemens confirms that QC and patient samples under-recovers MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Siemens sent an Urgent Medical Device Dorrection letter, dated August 2014, to all affected consignees. The letter identified the product, the problem, and the actions to be taken. Consignees were instructed if they still have an active calibration of MG with this lot of CHEM 1 CAL, they should immediately calibrate MG with an alternate lot of CHEM 1 CAL. Consigness were instructed to retain the letter with their laboratory records, and to forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Customer Care Center - Techhnical Solutions or their local Siemens technical support representative.
- Quantity in commerce
- 9769 Total (6000 US)
- Distribution
- Worldwide Distribution - USA (Nationwide) and Internationally to Australia Austria Belgium Canada Czech Republic Finland France Germany Great Britain Italy Japan Netherlands New Zealand Norway Portugal Slovak Republic Slovenia South Korea Spain Switzerland
- Product codes and lots
- lot 3GM081, exp. 07-01-2014
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2720-2014 | Class II | Terminated | Voluntary: Firm initiated |