Z-0009-2015Class II

eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN...

Gendex Corp. / initiated 2014-08-12 / Terminated / root cause: manufacturing process

Gendex Corp began a recall of eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach... on . FDA classified it as Class II and gives the reason as "Two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0009-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gendex Corp.
Product
eXpertDC intraoral x-ray, models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm & Tubehead, 765DC Product Usage: intraoral x-ray
Product code
EHD Unit, X-Ray, Extraoral With Timer
Reason for recall
Two (2) failures of the spring link components within the articulated arm assembly. The spring links broke during normal assembly process for the articulated arm during first cycle of the arm, prior to the 14lb tube head (full load) being installed onto the assembled articulated arm.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Gendex sent a Recall Notification Letter, dated August 12, 2014 and a second Communication Notification letter on September 22, 2014 to all affected customers. The letter identified the affected product, problem, and actions to be taken. For questions contact Gendex Customer Service at 1-800-323-8029.
Quantity in commerce
651
Distribution
Worldwide Distribution - US Nationwide and the countries of Europe, Canada, Australia, China, Russia, Taiwan, Singapore, and Thailand.
Product codes and lots
Models: eXpert DC, NA, 75 Inch reach; eXpert DC, NA, 65 Inch reach; eXpert DC, CN, 55 Inch reach; eXpert DC, WW, 75 Inch reach; eXpert DC, WW, 65 Inch reach; eXpert DC, WW, 55 Inch reach; OEM Arm Assembly w/o cables; Assy, Articulated Arm & Tubehead, eDC; and Assy, Articulated Arm & Tubehead, 765DC; with multiple part numbers and serial numbers: see distribution list for these details.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0009-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0009-2015Class IITerminatedVoluntary: Firm initiated