Z-2637-2014Class II

ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software...

Data Innovations, LLC / initiated 2014-08-15 / Terminated / root cause: software

Data Innovations, LLC began a recall of ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between... on . FDA classified it as Class II and gives the reason as "Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2637-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Data Innovations, LLC
Product
ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between in-vitro diagnostic instruments (IVDs) and Laboratory Information Systems (LIS) aiding the flow of laboratory information, including test orders and results, specimen information, and patient information.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.
Root cause tag
software
FDA root cause
Software design
Action
Data Innovations LLC sent an Important Medical Device Information letter on August 15, 2014, to consignees for the JResultNet using ProVue Instrument Driver JResultNet (provue.jrm) v. 1.5.3 and below. Potential issues with the ProVue driver have been identified which may cause data to be mis-associated with patient results under a given set of circumstances. The user on the ProVue transmits the same results twice back to back by clicking the Send to Host button twice. The ProVue sends two ASTM messages back to back before ending transmission (sending an EOT character) The solution was to disconnect JResultNet from the ProVue instrument and apply the patch to driver provue.jrm when it is available. If you have further questions regarding this notice, please feel free to contact us at [email protected] or 239-931-6004 Toll-Free (from the US and Canada): 1-800-332-0499.
Quantity in commerce
84
Distribution
Nationwide Distribution including AL, AZ, CA, CT, DC, FL, GA, IN, LA, MI, MO, NC, NH, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, and WA.
Product codes and lots
Version #s: 1.0.0 Date Released 6/17/2004, 1.0.1 Date Released 8/18/2004, 1.1.1 Date Released 9/9/2004, 1.2.0 Date Released 10/11/2004, 1.3.0 Date Released 11/1/2004, 1.3.1 Date Released 1/20/2005, 1.3.2 Date Released 2/18/2005, 1.4.0 Date Released 3/30/2005, 1.4.1 Date Released 8/31/2005, 1.4.2 Date Released 9/12/2005, 1.4.2 Date Released 10/5/2005, 1.4.3 Date Released 2/8/2006, 1.4.4 Date Released 2/20/2006, 1.4.5 Date Released 3/3/2006, 1.4.5 Date Released 3/23/2006, 1.4.5 Date Released 5/17/2006, 1.5.0 Date Released 6/26/2007, 1.5.1 Date Released 9/17/2007, 1.5.1 Date Released 12/13/2007, 1.5.1 Date Released 11/7/2008, 1.5.2 Date Released 4/13/2009, 1.5.3 Date Released 1/12/2011.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2637-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2637-2014Class IITerminatedVoluntary: Firm initiated