Z-2620-2014Class I

STAT 2 Flow Controller; 80" (203 cm) Appx 14.8 mL, 60 Drop Primary Set w/2 Clave, STAT 2 Flow Controller...

Abbott Medical / initiated 2014-08-19 / Terminated / root cause: manufacturing process

ICU Medical, Inc. began a recall of STAT 2 Flow Controller; 80" (203 cm) Appx 14.8 mL, 60 Drop Primary Set w/2 Clave, STAT 2 Flow Controller, Rotating Luer, Item No. 011-C9802. Product Usage: IV... on . FDA classified it as Class I and gives the reason as "ICU is recalling the ConMed Stat 2 flow controller because it may deliver a significantly higher flow rate than was intended". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2620-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
STAT 2 Flow Controller; 80" (203 cm) Appx 14.8 mL, 60 Drop Primary Set w/2 Clave, STAT 2 Flow Controller, Rotating Luer, Item No. 011-C9802. Product Usage: IV Administration sets are intended for the infusion and withdrawal of fluids. The devices are provided sterile.
Product code
FMG Stopcock, I.V. Set
Reason for recall
ICU is recalling the ConMed Stat 2 flow controller because it may deliver a significantly higher flow rate than was intended.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
ICU Medical sent an Urgent Medical Device Recall letter dated August 19, 2014 to all affected customers. The letter identified the affected product, problems identified and the actions to be taken. Customers were instructed to complete the recall response form and fax to ICU Medical at (801) 264-1755 or by email: [email protected]. Customers with questions or require assistance relating to this recall were instructed to contact ICU Customer Service Monday through Friday between the hours of 8:30 AM and 4:00 PM Pacific time: 866-829-9025 and select option 8, or e-mail the following address: [email protected]. For questions regarding this recall call 801-264-1732.
Quantity in commerce
1250 units
Distribution
Worldwide Distribution - US (nationwide) in the state of WA, OR and the country of South Africa.
Product codes and lots
Lot # 2768417

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2620-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2619-2014Class ITerminatedVoluntary: Firm initiated
Z-2620-2014Class ITerminatedVoluntary: Firm initiated