510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0974-2015Class II

TrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar Trypticase Soy Agar with 5% Sheep...

Becton Dickinson & Co / initiated 2014-06-06 / Terminated / root cause: manufacturing process

Becton Dickinson & Co. began a recall of TrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar Trypticase Soy Agar with 5% Sheep Blood is used for the growth of fastidious... on . FDA classified it as Class II and gives the reason as "A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0974-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
TrypticaseTM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar Trypticase Soy Agar with 5% Sheep Blood is used for the growth of fastidious organisms and for the visualization of hemolytic reactions. MacConkey II Agar is a selective and differential medium for the detection of coliform organisms and enteric pathogens.
Product code
JSI Culture Media, Selective And Differential
Reason for recall
A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as 1-3 CFUs per plate when present.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
BD Diagnostic Systems sent a Field Correction Notice dated August 2014 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to continue to follow all surveillance requirements of their laboratory upon confirmation of Listeria monocytogenes. Laboratory staff should notify their respective state health department. Customers with questions should contact BD Technical Services at 800-638-8663. For questions regarding this recall call 410-316-4000.
Quantity in commerce
7,617 cases (100 plates/case)
Distribution
Worldwide Distribution - US nationwide including Puerto Rico, Belgium, Canada, Philippines, Singapore, and Taiwan.
Product codes and lots
Catalogue # 221291 with Lot #4050354,  4086399,  4099206 and 4143653; Catalogue #221261 with Lot # 4022158 4086443 4092199

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0974-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0974-2015Class IITerminatedVoluntary: Firm initiated
Z-0975-2015Class IITerminatedVoluntary: Firm initiated
Z-0976-2015Class IITerminatedVoluntary: Firm initiated