Z-2683-2014Class II

The Verigene Gram-Positive Blood Culture Nucleic Acid Test (BC-UP) performed using the sample-to-result...

Luminex Corporation / initiated 2013-10-10 / Terminated / root cause: manufacturing process

Nanosphere, Inc. began a recall of The Verigene Gram-Positive Blood Culture Nucleic Acid Test (BC-UP) performed using the sample-to-result Verigene System is a qualitative, multiplexed in vitro... on . FDA classified it as Class II and gives the reason as "Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2683-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Luminex Corporation
Product
The Verigene Gram-Positive Blood Culture Nucleic Acid Test (BC-UP) performed using the sample-to-result Verigene System is a qualitative, multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause bloodstream infection (BSI). BC-UP is performed directly on blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria. BC-UP detects and identifies the following bacterial genera and species: Staphylococcus spp., Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus lugdunensis, Enterococcus faecalis, Enterococcus faecium, Streptococcus spp., Streptococcus pneumoniae, Streptococcus pyogenes, Streptococcus agalactiae, Streptococcus anginosus, group and Listeria spp. In addition, BC-UP detects the mnecA resistance marker, inferring mecA-mediated, methicillin resistance, and the vanA and vanB resistance markers, inferring vanA/vanB-mediated vancomycin resistance. In mixed growth, BC-UP does not specifically attribute van-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA-mediated methicillin resistance to either S. aureus or S. epidermidis. BC-UP is indicated for use in conjunction with other clinical and laboratory findings to, aid in the diagnosis of bacterial bloodstream infections; however, is not to be used to, monitor these infections. Sub-culturing of positive blood cultures is necessary to recover, organisms for susceptibility testing, identification of organisms not detected by BC-UP, differentiation of mixed growth, association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.
Product code
PAM Gram-Positive Bacteria And Their Resistance Markers
Reason for recall
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.
Quantity in commerce
4,440 units
Distribution
Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.
Product codes and lots
Product Number: 20-009-018; Lot Numbers and Expiration Dates: 1) Lot: 090613018B, Expiration Date: 03/05/14; 2) Lot: 091413018B, Expiration Date: 03/13/14; 3) Lot: 091713018C, Expiration Date: 03/16/14; 4) Lot: 091813018B, Expiration Date: 03/17/14; 5) Lot: 092113018C, Expiration Date: 03/20/14; 6) Lot: 100213018B, Expiration Date: 03/31/14

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2683-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2681-2014Class IITerminatedVoluntary: Firm initiated
Z-2682-2014Class IITerminatedVoluntary: Firm initiated
Z-2683-2014Class IITerminatedVoluntary: Firm initiated
Z-2684-2014Class IITerminatedVoluntary: Firm initiated
Z-2685-2014Class IITerminatedVoluntary: Firm initiated